cardiac_mri:protocols:structural
Structural
<callout type=“info” icon=“true”> Identical locations must be used for each sequence if possible. Always used the “Copy image location” tool when setting up subsequent sequences </callout>
- Localizer
- 3 plane SSFP
- Repeat in ref if needed to move heart to isocenter, then repeat in ref and delete all but final loc
- Axial SSFP
- include the entire aortic arch to diaphragm
- Axial HASTE
- Free breathing, include the entire aortic arch to diaphragm
- Pseudo-SA
- Perpindicular to long axis Loc
- 4ch cine SSFP
- 3PT at pseudo-SA RV free wall, LV free wall, apex from loc
- Avoid the aortic valve
- 2ch cine SSFP
- 3PT at pseudo-SA LV anterior wall, LV inferior wall, from previous CINE
- 3ch cine SSFP
- 3PT at pseudo-SA aortic valve, pseudo-SA opposite point to AoV, apex from previous CINE
- SA cine SSFP Stack
- Perpindicular to 4 chamber. Stack at 10 mm intervals to cover the entire left ventricle
cardiac_mri/protocols/structural.txt · Last modified: by orhan